EXCOR Active Driving System for the EXCOR Pediatric VAD IDE Study

The safety and scientific validity of this study is the responsibility of the study sponsor and investigators. Listing a study does not mean it has been evaluated by the U.S. Federal Government. Know the risks and potential benefits of clinical studies and talk to your health care provider before participating. Read our disclaimer for details.
 
ClinicalTrials.gov Identifier: NCT05610787
Recruitment Status : Recruiting
First Posted : November 9, 2022
Last Update Posted : February 8, 2023
Sponsor:
Information provided by (Responsible Party):
Berlin Heart, Inc

Tracking Information
First Submitted Date  ICMJE September 14, 2022
First Posted Date  ICMJE November 9, 2022
Last Update Posted Date February 8, 2023
Actual Study Start Date  ICMJE November 14, 2022
Estimated Primary Completion Date September 2023   (Final data collection date for primary outcome measure)
Current Primary Outcome Measures  ICMJE
 (submitted: November 8, 2022)
  • Rate of Device Malfunction adverse events [ Time Frame: Up to 90 days ]
    A Device Malfunction per protocol definition, and adjudicated by the Clinical Events Committee.
  • Patient Outcome [ Time Frame: Up to 90 days ]
    Patient outcomes will be summarized as the proportion of subjects experiencing each outcome and the overall successful outcome. Successful outcome is defined as:
    • survival to recovery/successful weaning -or-
    • survival to explantation not attributed to device malfunction -or-
    • survival on EXCOR® Pediatric at 90 days post-implant
    Patients who are removed from support followed by death or for escalation to other therapy (such as another VAD or ECMO) due to major device malfunction attributed to the Active Driving System will be considered failures.
  • Serious Adverse Events [ Time Frame: Up to 90 days ]
    Serious Adverse Event rates will be calculated as the rate per 100 patient-months for each individual event as defined in protocol.
Original Primary Outcome Measures  ICMJE Same as current
Change History
Current Secondary Outcome Measures  ICMJE Not Provided
Original Secondary Outcome Measures  ICMJE Not Provided
Current Other Pre-specified Outcome Measures Not Provided
Original Other Pre-specified Outcome Measures Not Provided
 
Descriptive Information
Brief Title  ICMJE EXCOR Active Driving System for the EXCOR Pediatric VAD IDE Study
Official Title  ICMJE EXCOR Active Driving System for the EXCOR Pediatric VAD IDE Study
Brief Summary

The purpose of this study is to evaluate the device performance and monitor the safety and effectiveness of the Berlin Heart EXCOR Active Driving System while being used with the approved EXCOR Pediatric Ventricular Assist Device.

EXCOR Active Driving System is intended for use with the approved EXCOR Pediatric VAD.

The EXCOR Pediatric VAD is intended to provide mechanical circulatory support as a bridge to cardiac transplantation for pediatric patients. Pediatric candidates with severe isolated left ventricular or biventricular dysfunction who are candidates for cardiac transplant and require circulatory support may be treated using the EXCOR Pediatric.

EXCOR Active is intended for use in a clinical setting. EXCOR Active can be used in any kind of hospital unit (e.g. OR, ICU, intermediate care unit or general care unit). The driving unit may be moved between clinical units using the caddy or baby buggy; however, a patient must always be accompanied by a person trained in the use of the manual pump and emergency procedures during transport in the event of an emergency.

The driving unit can be transported during operation.

Detailed Description

This is a prospective study of subjects implanted with the EXCOR Pediatric VAD supported by the EXCOR Active Driving System and patients recently implanted with EXCOR Pediatric supported on the Ikus Driving Unit as reported in the ACTION Registry database.

Children who are transplant eligible in need of mechanical circulatory support and are supported with the EXCOR® Pediatric VAD will be included in this study. All patients must meet the indications and contraindication as indicated in the device labelling. There are no known additional indications or contraindications associated with the EXCOR® Active Driving System beyond those already established in the EXCOR® Pediatric VAD labeling.

The objective of this study is to evaluate device performance and monitor safety and effectiveness of the Berlin Heart EXCOR® Active Driving System while being used with the approved EXCOR® Pediatric Ventricular Assist Device.

Study Type  ICMJE Interventional
Study Phase  ICMJE Not Applicable
Study Design  ICMJE Allocation: Non-Randomized
Intervention Model: Parallel Assignment
Intervention Model Description:
Open Label, Single Arm, Prospective, Non-Randomized
Masking: None (Open Label)
Primary Purpose: Treatment
Condition  ICMJE
  • Heart Failure
  • Transplant; Failure, Heart
  • Congenital Heart Disease
Intervention  ICMJE Device: EXCOR Active Driving System for Pediatric VAD
The EXCOR Active Driving System - will be used with the EXCOR Pediatric VAD.
Study Arms  ICMJE
  • Experimental: Active Driver Primary Arm
    Up to 40 patients whom receive the Active Driver from implant.
    Intervention: Device: EXCOR Active Driving System for Pediatric VAD
  • No Intervention: IKUS Comparator
    All patients whom receive the (FDA approved) IKUS Driver from implant and are entered into the ACTION Registry.
Publications * Not Provided

*   Includes publications given by the data provider as well as publications identified by ClinicalTrials.gov Identifier (NCT Number) in Medline.
 
Recruitment Information
Recruitment Status  ICMJE Recruiting
Estimated Enrollment  ICMJE
 (submitted: November 8, 2022)
40
Original Estimated Enrollment  ICMJE Same as current
Estimated Study Completion Date  ICMJE January 2024
Estimated Primary Completion Date September 2023   (Final data collection date for primary outcome measure)
Eligibility Criteria  ICMJE

Inclusion Criteria:

  • Patient meets all indications as specified in the current version of the Instructions for Use (IFU) of the EXCOR® Pediatric VAD system as shown below:

"EXCOR is intended to provide mechanical circulatory support as a bridge to cardiac transplantation for pediatric patients. Pediatric patients with severe isolated left ventricular or biventricular dysfunction who are candidates for cardiac transplant and require circulatory support may be treated using EXCOR."

• Patient and/or legal representative has signed the study informed consent form.

Exclusion Criteria:

• Patient has any contraindication as specified in the current version of the IFU of the EXCOR® Pediatric VAD system as shown below:

"Patients unable to tolerate systemic anticoagulation therapy should not be implanted with EXCOR components.

Magnetic Resonance Imaging (MRI) is contraindicated in patients after being implanted with EXCOR.

Patients with aortic valve regurgitation that is more than moderate that cannot be re- paired at the time of implantation should not be implanted with EXCOR. If repair of the aortic valve regurgitation requires surgical closure of the aortic valve, EXCOR should not be implanted. EXCOR is not intended to be used as a total artificial heart and should not be used in this configuration."

  • Patient and/or legal representative has not given written consent to participate in the study (non-consent).
  • Females of childbearing age who are not on contraceptives or surgically sterile or pregnant.
Sex/Gender  ICMJE
Sexes Eligible for Study: All
Ages  ICMJE up to 21 Years   (Child, Adult)
Accepts Healthy Volunteers  ICMJE No
Contacts  ICMJE
Contact: Mary Beth Kepler 734-709-2080 mkepler@berlinheartinc.com
Contact: Robert Kroslowitz 281-863-9700 rkroslowitz@berlinheartinc.com
Listed Location Countries  ICMJE United States
Removed Location Countries  
 
Administrative Information
NCT Number  ICMJE NCT05610787
Other Study ID Numbers  ICMJE G200252
Has Data Monitoring Committee Yes
U.S. FDA-regulated Product
Studies a U.S. FDA-regulated Drug Product: No
Studies a U.S. FDA-regulated Device Product: Yes
IPD Sharing Statement  ICMJE
Plan to Share IPD: No
Current Responsible Party Berlin Heart, Inc
Original Responsible Party Same as current
Current Study Sponsor  ICMJE Berlin Heart, Inc
Original Study Sponsor  ICMJE Same as current
Collaborators  ICMJE Not Provided
Investigators  ICMJE Not Provided
PRS Account Berlin Heart, Inc
Verification Date December 2022

ICMJE     Data element required by the International Committee of Medical Journal Editors and the World Health Organization ICTRP